Model | Measurement Notes (below) |
Units | Install Date |
Removal Date |
Current SN |
Current CF |
---|---|---|---|---|---|---|
LI-200 | Global Horizontal | W/m2 | 02/03/2009 | 08/24/2013 | PY61283 | -10.44 W/m2/µA |
08/26/2013 | - | PY29660 | -10.95 W/m2/µA | |||
Measured using a LICOR LI-200 Pyranometer mounted on an Irradiance Inc. Rotating Shadowband Radiometer (RSR). RSR mounted on the ground and the LI-200 sensor height is approx. 7 feet above ground level. The uncorrected value is for testing and troubleshooting purposes only. Voltage is measured across a 100 Ohm precision resistor in parallel to sensor output. | ||||||
LI-200 | Diffuse Horizontal | W/m2 | 02/03/2009 | 08/24/2013 | PY61283 | -10.44 W/m2/µA |
08/26/2013 | - | PY29660 | -10.95 W/m2/µA | |||
Measured using the above LI-200, when RSR band rotates every 30-seconds and blocks the sun. The uncorrected value is for testing and troubleshooting purposes only. Voltage is measured across a 100 Ohm precision resistor in parallel to sensor output. | ||||||
(calc) | Direct Normal | W/m2 | 02/03/2009 | 08/26/2013 | - | FD = 1.03378 |
08/26/2013 | - | - | FD = 1.0 | |||
Calculated using Global and Diffuse measurement. The uncorrected value is for testing and troubleshooting purposes only. | ||||||
LI-200 | Global Horizontal (secondary) | W/m2 | 02/25/2009 | 08/24/2013 | PY61743 | -10.36 W/m2/µA |
08/26/2013 | - | PY29660 | -11.98 W/m2/µA | |||
Measured using a LICOR LI-200 Pyranometer mounted next to the primary LI-200, but outside of the rotating shading band. Voltage is measured across a 100 Ohm precision resistor in parallel to sensor output. | ||||||
PSP | Global Horizontal | W/m2 | 08/18/2011 | 08/24/2013 | 27516F3 | 126.67 W/m2/mV |
PSP | 08/26/2013 | 01/31/2015 | 21231F3 | 109.65 W/m2/mV | ||
SPP | 01/31/2015 | - | 18851F3 | 117.92 W/m2/mV | ||
Total Hemispheric shortwave irradiance as measured by an Eppley Laboratory, Inc. Thermopile Pyranometer with calibration factor traceable to the World Radiometric Reference (WRR). | ||||||
CS215 | Dry Bulb Temperature and Relative Humidity | oC % |
02/03/2009 | - | - | N/A |
Outside air temperature and relative humidity inside in a naturally aspirated radiation shield mounted approx. 5 feet above ground level. | ||||||
CS106 | Station Pressure | mBar | 02/03/2009 | - | - | 0.24 mBar/mV+500mBar |
The station pressure is measured by a Vaisala sensor mounted inside weatherproof housing. This sensor is only measured every 15-minutes to conserve battery power. | ||||||
(calc) | Zenith Angle | o | 02/03/2009 | - | - | - |
Solar zenith angle calculated by RSR datalogger program. | ||||||
- | RSR Battery Voltage | V | 02/25/2009 | - | - | - |
The RSR Battery powers the datalogger and the shadowband, it is charged by a PV panel. | ||||||
- | CR1000 Temperature | oC | 02/25/2009 | - | - | - |
Temperature of CR1000 datalogger panel. Logger is mounted inside weatherproof enclosure. Note: The logger was upgraded from a CR800 to a CR1000 on 8/10/2010. | ||||||
Notes:
|