1ATIATI06/08/201400:00:00The SRRL ATI has been malfunctioning on various days, symptoms are direct near zero and global = diffuse. Data should not be used.
2ATIATI01/16/200610:49:0001/30/200611:19:00The SRRL ATI RSP head unit (including the LICOR PY21094) was swapped out with a head unit to be calibrated (PY17942). The calibration factor in the datalogger program was not changed, therefore data is out of calibration during this time frame. CF (in logger) = 114.68; CF (PY17942) = 109.76 (after calibraiton).
3ATIATI03/27/200106:50:0003/27/200123:59:00The ATI shadowband stopped rotating at approx 7:00 MST, data maybe missing and/or bogus during this day.
4ATIATI06/13/200007:50:0006/14/200015:20:00The ATI was removed from the old deck and installed on the new SRRL instrument deck. Data during this time frame is missing.