R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Global TUVR | 05/08/2017 | 13:30:00 | 05/09/2017 | 10:00:00 | Instrument is reading bogus after heavy hail storm at SRRL destroyed signal cable. Cable was replaced. |
2 | BMS | 33444 | Global TUVR | 02/17/2014 | 19:00:00 | 02/18/2014 | 11:19:00 | TUVR signal is low and noisy due to a loose screw in the patch panel. Screw was tightened and instrument returned to service at 11:19 MST. |
3 | BMS | 33444 | Global TUVR | 02/07/2006 | 10:30:00 | 02/14/2006 | 10:20:00 | TUVR signal is low, erratic, and/or negative. Unit was shorted to reset. |
4 | BMS | 33444 | Global TUVR | 07/10/2005 | 00:00:00 | 07/18/2005 | 12:21:00 | TUVR signal is low, erratic, and/or negative. Unit was shorted to reset. Also during a channel short test it was discovered that a semi-loose screw in the patch panel was causing intermittent problems. |
5 | BMS | 33444 | Global TUVR | 07/07/2005 | 00:00:00 | 07/08/2005 | 12:20:00 | TUVR signal is low, erratic, and/or negative. Unit was shorted to reset. |
6 | BMS | 33444 | Global TUVR | 05/13/2005 | 12:55:00 | 05/23/2005 | 12:00:00 | TUVR signal is low, erratic, and/or negative. Unit was shorted to reset. |
7 | BMS | 33444 | Global TUVR | 05/10/2005 | 00:00:00 | 05/10/2005 | 14:10:00 | TUVR signal is low, erratic, and/or negative. Unit was shorted to reset. |
8 | BMS | 33444 | Global TUVR | 03/11/2005 | 09:30:00 | 04/26/2005 | 10:20:00 | TUVR signal is low, erratic, and/or negative. |
9 | BMS | 33444 | Global TUVR | 03/09/2005 | 08:30:00 | 03/09/2005 | 13:00:00 | TUVR signal was low and erratic during this time. |
10 | BMS | 33444 | Global TUVR | 02/09/2005 | 00:00:00 | 02/09/2005 | 23:59:00 | TUVR signal was low and erratic for one day. |
11 | BMS | 33444 | Global TUVR | 01/05/2005 | 15:38:00 | 01/10/2005 | 16:01:00 | TUVR has failed again, output is negative. Unit was disconnected and shorted and now seems to be working again. |
12 | BMS | 33444 | Global TUVR | 12/20/2004 | 00:00:00 | 12/20/2004 | 23:59:00 | TUVR signal was low and somewhat erratic for one day. |
13 | BMS | | Global TUVR | 03/12/2004 | 00:00:00 | 11/18/2004 | 11:13:00 | The Global TUVR is experiencing intermitten signal problems, some days its good others its bad. The instrument was replaced with TUVR s/n 33444, which seemed to fixed the problem. |
14 | BMS | | Global TUVR | 03/03/2004 | 10:30:00 | 03/10/2004 | 10:43:00 | The Global TUVR is experiencing intermitten signal problems. On 3-10-2004, the TUVR sensor was shorted from 10:11 to 10:42 MST in an effort to reset the detector. |
15 | BMS | | Global TUVR | 02/03/2003 | 00:00:00 | 04/16/2003 | 08:45:00 | The BMS Global TUVR is having intermitten problems again. The signal has been low or erratic since Feb 3. |
16 | BMS | | Global TUVR | 01/17/2003 | 13:00:00 | 01/27/2003 | 10:30:00 | The Global TUVR (S/N 33444) is experiencing intermitten signal problems, signal is low or erratic. The instrument was replaced with a factory refurbished TUVR s/n 18180, which seems to have fixed the problem. |
17 | BMS | | Global TUVR | 03/26/2002 | 13:30:00 | 05/09/2002 | 08:47:00 | The Global TUVR is experiencing intermitten signal problems, some days its good others its bad. The instrument was replaced with new TUVR s/n 33444, which fixed the problem. |
18 | BMS | | Global TUVR | 08/27/2001 | 13:00:00 | 01/16/2002 | 09:43:00 | Intermitten signal problems, some days it's good, other days the signal is low. Due to old signal cable from the instrument to the J-Box, the cable was replaced. |