MIDC/Data Quality Statements
Total Records: 18, displaying 1-18

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSGlobal TUVR05/08/201713:30:0005/09/201710:00:00Instrument is reading bogus after heavy hail storm at SRRL destroyed signal cable. Cable was replaced.
2BMS33444Global TUVR02/17/201419:00:0002/18/201411:19:00TUVR signal is low and noisy due to a loose screw in the patch panel. Screw was tightened and instrument returned to service at 11:19 MST.
3BMS33444Global TUVR02/07/200610:30:0002/14/200610:20:00TUVR signal is low, erratic, and/or negative. Unit was shorted to reset.
4BMS33444Global TUVR07/10/200500:00:0007/18/200512:21:00TUVR signal is low, erratic, and/or negative. Unit was shorted to reset. Also during a channel short test it was discovered that a semi-loose screw in the patch panel was causing intermittent problems.
5BMS33444Global TUVR07/07/200500:00:0007/08/200512:20:00TUVR signal is low, erratic, and/or negative. Unit was shorted to reset.
6BMS33444Global TUVR05/13/200512:55:0005/23/200512:00:00TUVR signal is low, erratic, and/or negative. Unit was shorted to reset.
7BMS33444Global TUVR05/10/200500:00:0005/10/200514:10:00TUVR signal is low, erratic, and/or negative. Unit was shorted to reset.
8BMS33444Global TUVR03/11/200509:30:0004/26/200510:20:00TUVR signal is low, erratic, and/or negative.
9BMS33444Global TUVR03/09/200508:30:0003/09/200513:00:00TUVR signal was low and erratic during this time.
10BMS33444Global TUVR02/09/200500:00:0002/09/200523:59:00TUVR signal was low and erratic for one day.
11BMS33444Global TUVR01/05/200515:38:0001/10/200516:01:00TUVR has failed again, output is negative. Unit was disconnected and shorted and now seems to be working again.
12BMS33444Global TUVR12/20/200400:00:0012/20/200423:59:00TUVR signal was low and somewhat erratic for one day.
13BMSGlobal TUVR03/12/200400:00:0011/18/200411:13:00The Global TUVR is experiencing intermitten signal problems, some days its good others its bad. The instrument was replaced with TUVR s/n 33444, which seemed to fixed the problem.
14BMSGlobal TUVR03/03/200410:30:0003/10/200410:43:00The Global TUVR is experiencing intermitten signal problems. On 3-10-2004, the TUVR sensor was shorted from 10:11 to 10:42 MST in an effort to reset the detector.
15BMSGlobal TUVR02/03/200300:00:0004/16/200308:45:00The BMS Global TUVR is having intermitten problems again. The signal has been low or erratic since Feb 3.
16BMSGlobal TUVR01/17/200313:00:0001/27/200310:30:00The Global TUVR (S/N 33444) is experiencing intermitten signal problems, signal is low or erratic. The instrument was replaced with a factory refurbished TUVR s/n 18180, which seems to have fixed the problem.
17BMSGlobal TUVR03/26/200213:30:0005/09/200208:47:00The Global TUVR is experiencing intermitten signal problems, some days its good others its bad. The instrument was replaced with new TUVR s/n 33444, which fixed the problem.
18BMSGlobal TUVR08/27/200113:00:0001/16/200209:43:00Intermitten signal problems, some days it's good, other days the signal is low. Due to old signal cable from the instrument to the J-Box, the cable was replaced.