MIDC/Data Quality Statements
Total Records: 2, displaying 1-2

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMS2703Global PYR-P08/14/200700:00:0010/20/200810:37:00Instrument is reading 10-20% low for some unknown reason. Signal was verified to be the same on deck and at logger. Instrument may be degrading, however some erratic behavior was also observed on Sept. 3 & 4 when the instrument had several spikes of irradiance levels up to 2500 W/m^2 for periods up to about 15-minutes, then back to "normal".
2BMS2703Global PYR-P03/25/200600:00:0005/31/200607:48:00Instrument was reading 15-20% high for some unknown reason. Signal was verified to be the same on deck and at logger. Instrument was calibrated in BORCAL 2006-01 and new factor was entered into logger at 7:48 MST on May 31.