MIDC/Data Quality Statements
Total Records: 7, displaying 1-7

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1OTF700G4S07125.02Global 40-South MS-70011/10/201602/15/201811:23:00On 2-15-2018 the MS-700 was received for calibration and was reading 28% low across all wavelengths. It is unknown when the problem started or if it was gradual or sudden.
2OTF700G4S07125.02Global 40-South MS-70010/07/201206:56:0010/08/201207:38:00The MS-700 is putting out duplicate spectra during this time. This data is not usable and has been removed. Problem is most due to power glitch to NREL network infrastructure on 10/7/2012 at 6:55:15 MST which caused manufacture data collection software to produce duplicate readings until the software was manually restarted on 10/8/2012 at 7:39 MST. On 10/18/2012 the manufacture software was replaced with NREL developed software which will completely eliminate this particular problem.
3OTF700G4S07125.02Global 40-South MS-70009/15/201209:35:0009/17/201208:09:00The MS-700 is putting out duplicate spectra during this time. This data is not usable and has been removed. Problem is most likely due to NREL network infrastructure maintenance which caused manufacture data collection software to produce duplicate readings until the software was manually restarted at 8:10 MST.
4OTF700G4S07125.02Global 40-South MS-70007/06/201204:30:0007/06/201207:41:00The MS-700 is putting out duplicate spectra during this time. This data is not usable and has been removed. Problem is due to power glitch at NREL at 4:05 MST that resulted in loss of power to ethernet switch at OTF that affected ethernet communication which caused manufacture data collection software to produce duplicate readings until the software was manually restarted at 7:42 MST.
5OTF700G4S07125.02Global 40-South MS-70006/22/201204:31:0006/22/201207:30:00The MS-700 is putting out duplicate spectra during this time. This data is not usable and has been removed. Problem is due to power glitch at NREL at 1:31 MST that resulted in loss of power to ethernet switch at OTF that affected ethernet communication which caused manufacture data collection software to produce duplicate readings until the software was manually restarted at 7:30 MST.
6OTF700G4S07125.02Global 40-South MS-70005/19/201204:30:0005/26/201207:11:00The MS-700 is putting out duplicate spectra during this time. This data is not usable and has been removed. Problem is most likely the result of power glitch at OTF that affected ethernet communication which caused manufacture data collection software to produce duplicate readings until the software was manually restarted.
7OTF700G4S07125.02Global 40-South MS-70002/22/201207:34:0002/29/201212:02:00The MS-700 is putting out duplicate spectra during this time. This data is not usable and has been removed. Problem is most likely the result of power glitch at OTF that affected ethernet communication which caused manufacture data collection software to produce duplicate readings until the software was manually restarted.