R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | S16059.01 | Global Horizontal MS-712 | 09/03/2021 | 07:50:00 | 09/02/2022 | | Based on cosine response testing conducted on 15-Sept-2021 and 22-Sept-2021 it has been determined that the MS-712 morning data is up to 8% low and afternoon data is up to 8% high. Instrument was swapped with another serial number that had been repaired for cosine response on 9/2/2022. |
2 | BMS | S18070.01 | Global Horizontal MS-712 | 07/14/2021 | 11:11:00 | 09/03/2021 | 07:50:00 | Based on cosine response testing conducted on 12-Aug-2021 it has been determined that the MS-712 afternoon data is up to 20% high. Instrument was swapped with another serial number on 9/3/2021. |
3 | BMS | S13099.11 | Global Horizontal MS-712 | 03/13/2014 | 00:00:00 | 07/14/2021 | 11:11:00 | Based on cosine response testing conducted on 24-May-2021 and 27-May-2021 it has been determined that the MS-712 afternoon data is up to 20% low. Instrument was swapped with another serial number on 7/14/2021. |