R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Global 40-South CUV4 | 05/13/2022 | 11:45:00 | 05/13/2022 | 12:30:00 | The 40-South UV sensor was moved to horizontal during this approximate time so that it could be calibrated. |
2 | BMS | | Global CUV4 | 05/10/2015 | 05:00:00 | 05/11/2015 | 13:56:00 | Water infiltrated the outdoor bendix connector on the j-box. Opened and used compressed air to dry at 13:56. Cable and bendix connector will be replaced in near future. |
3 | BMS | | Global CUV4 | 04/29/2015 | 06:20:00 | 04/29/2015 | 11:26:00 | Water infiltrated the outdoor bendix connector on the j-box. Opened and used compressed air to dry at 11:26. |
4 | BMS | | Global CUV4 | 12/21/2014 | 22:00:00 | 12/25/2014 | 18:00:00 | Water infiltrated the outdoor bendix connector on the j-box, because signal cable was too short for temporary j-box that was on the ground. Water dripped directly down cable into connector which is causing bogus readings. Connector was dried on Dec 23 at 12:40, however signal may be reading very slightly low until Dec 25 at 18:00, when data appears to have fully recovered (no more nighttime offset). |
5 | BMS | | Global CUV4 | 11/30/2010 | 16:00:00 | 11/30/2010 | 16:30:00 | Instrument was reading bogus during this time while the instrument was being re-mounted on a taller mount. |