R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSGlobal CMP22 Vent06/16/202205:05:0006/16/202205:49:00Preliminary BORCAL data from June 15 shows R@70=10.6 (AM) and R@70=10.3 (PM). Instrument bubble level was adjusted on June 16, instrument may be shaded at random during time range above during adjustment. On July 13, possible bogus data from during this time was corrected by substituting with Global CMP22-1.
2BMSGlobal CMP22 Vent06/10/202211:13:0007/06/202216:01:00New fan that was installed on June 10, was put in backwards. Fan orientation was corrected on July 6. Also substitution was not enabled in time, so data is bogus from 15:20 to 16:01 when fan was being corrected on July 6. On July 13, bogus data from 15:20 to 16:01 on July 6 was corrected by substituting with Global CMP22-1.
3BMSGlobal CMP22 Vent06/09/202213:08:0006/10/202211:12:00During this time the CMP22 Vent data was being automatically substituted with the CMP22-1 in the data logger. The bubble level replaced (the old one was dry) and the ventilator fan/filter replaced (the fan was a bit noisy and flow a bit low). The installation of the bubble still requires verification through BORCAL.
4BMSGlobal CMP22 Vent05/18/202013:03:0012/15/202119:31:00DATA CORRECTED on 12/17/2021: Irradiances were increased by 0.675% due to revised BORCAL report for all BMS shortwave radiometers calibrated in 2020 and 2021.