R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Global CM6b | 05/18/2020 | 13:03:00 | 12/15/2021 | 19:31:00 | DATA CORRECTED on 12/17/2021: Irradiances were increased by 0.675% due to revised BORCAL report for all BMS shortwave radiometers calibrated in 2020 and 2021. |
2 | BMS | 015189 | Global CM6b | 06/22/2011 | 09:00:00 | 06/24/2011 | 14:00:00 | Condensation has formed inside the CM6b dome near the top, which is causing lower irradiance readings around solar noon. The desiccant was replaced which seemed to dry out the dome after a couple days. |
3 | BMS | 015189 | Global CM6b | 06/15/2010 | 08:00:00 | 06/29/2010 | 16:00:00 | Condensation has formed inside the CM6b dome near the top, which is causing lower irradiance readings around solar noon. Instrument was opened, dried, and re-sealed between 15:30 to 15:45 MST on 6/29/2010. |
4 | BMS | 015189 | Global CM6b | 05/27/2009 | 00:00:00 | 06/05/2009 | 10:51:00 | Condensation has formed inside the CM6b dome near the top, which is causing lower irradiance readings around solar noon. Instrument was removed from service 6/1/2009 10:19 MST, dried, dome and desiccant o-rings replaced, and desiccant replaced. Returned to service 6/5/2009 10:51 MST. |