R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
---|---|---|---|---|---|---|---|---|
1 | BMS | Global 40-South PSP | 12/16/2002 | 14:43:00 | 12/17/2002 | 11:01:00 | The J-Box on the BMS Radiometer Tower was being replaced, data from this instrument is missing or bogus. | |
2 | BMS | Global 40-South PSP | 11/01/2000 | 12/03/2001 | 09:47:00 | Someone swapped the instrument without our knowledge, therefore the calibration factor was never updated in the logger program. The orignal instrument was a s/n 25766F3 the instrument was changed to a s/n 20715F3. The change likely occured around November 2000, the data was about 10% high. A calibration on the 20715F3 was performed and its calibration factor was entered into the logger on 12-3-2001. Start date is approximate. |