MIDC/Data Quality Statements
Total Records: 4, displaying 1-4

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSDirect MS-5605/18/202013:03:0012/15/202119:31:00DATA CORRECTED on 12/17/2021: Irradiances were increased by 0.675% due to revised BORCAL report for all BMS shortwave radiometers calibrated in 2020 and 2021.
2BMSP12022Direct MS-5605/24/201808:00:00The MS-56 has a large positive offset in the data. Channel was shorted on 5/28/2018 on instrument platform and it tested okay. Problem is therefore related to instrument or instrument cable.
3BMSDirect MS-5610/09/201714:50:0010/09/201716:00:00MS-56 and sNIP are out of alignment during this time. Data is low and should not be used.
4BMSP12022Direct MS-5604/27/201500:00:0005/11/201513:30:00The MS-56 has a large offset in the data. Problem started after rain showers. Instrument was removed from service on April 30 @ 14:45 MST for troubleshooting. Instrument was later returned to service on May 6 @ 16:03 MST (this time bypassing the AM25T solid state multiplexer), however due to insufficient sun the necessary shims could not be added to set alignment, data will read about 100 W/m^2 too low until the shims were added on May 11 @ 13:30 MST. Original problem is assumed to have been caused by the MS-56 high impedance (5 KOhm) in conjunction with the use of the solid state multiplexer.