R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Diffuse UVB-1 | 12/29/2007 | 00:00:00 | 01/03/2008 | 23:59:00 | Sci-Tec Tracker #2 had intermittent alignment problems. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. The AOCS was not pointing at the SUN. The affected times were: 12/29/07: 13:45-sunset; 12/31/07 @ 13:30 through 1/3/08. |
2 | BMS | | Diffuse UVB-1 | 11/28/2007 | 12:30:00 | 11/28/2007 | 15:00:00 | Sci-Tec Tracker #2 had intermittent alignment problems. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. The AOCS was not pointing at the SUN. |
3 | BMS | | Diffuse UVB-1 | 11/05/2007 | 00:00:00 | 11/11/2007 | 23:59:00 | Sci-Tec Tracker #2 had intermittent alignment problems. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. The AOCS was not pointing at the SUN. The affected times were: 11/5: sunrise to 8:45 MST; 11/7 through 11/11. |
4 | BMS | 921106 | Diffuse UVB-1 | 07/06/2007 | 17:35:00 | 07/09/2007 | 07:00:00 | The temperature controller was not running during this time, due to a tracker tripping the GFI that powers the UVB-1 TEC. |
5 | BMS | 921106 | Diffuse UVB-1 | 05/25/2006 | 10:50:00 | 05/25/2006 | 11:10:00 | Sci-Tec tracker #2 tracker was down due to power shutdown for installation of new SRRL equipment lift. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. |
6 | BMS | 921106 | Diffuse UVB-1 | 11/05/2005 | 06:37:00 | 11/07/2005 | 09:19:00 | Sci-Tec Tracker #2 was down due to a power outage on 11/4/2005 at 23:14 MST. Tracker failed to automatically reposition when power was restored on 11/5/2005 at 5:02 MST, due to failed internal battery. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. The AOCS was not pointing at the SUN. |
7 | BMS | 921106 | Diffuse UVB-1 | 07/23/2005 | 10:16:00 | 07/25/2005 | 07:07:00 | Sci-Tec Tracker #2 rotated the wrong direction and hung up on the signal cables, due to a scheduled NREL power outage on 7/23/2005 at 5:33 MST. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. |
8 | BMS | 921106 | Diffuse UVB-1 | 07/10/2005 | 05:56:00 | 07/11/2005 | 08:00:00 | Sci-Tec Tracker #2 rotated the wrong direction and hung up on the signal cables, due to a momenatary power loss. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. |
9 | BMS | | Diffuse UVB-1 | 10/29/2003 | 11:00:00 | 02/11/2004 | 00:00:00 | The Diffuse UVB-1 was removed from service and installed in the Global UVB-1 position. The original global instrument (S/N 921106) was returned to the factory, due to a residue that appeared inside the dome. A new dome was installed at the factory and the instrument was calibrated against the BMS Global UVB-1 on 2/10/2004 from 11:26 to 13:46 MST. This instrument (S/N 921106) is now installed in the Diffuse UVB-1 position. |