MIDC/AIM Calibration Database
Total Records: 4, displaying 1-4

R#Serial NumberCalibration FacilityCalibration DateDate for next CalibrationResponsivity [45]Calibration Factor [45]Instrument::MakeInstrument::ModelInstrument::DescriptionInstrument::ProgramInstrument::Factory CalibrationInstrument::Date of Factory CalibrationResponsivity Bins [2 deg]Responsivity Function CoefficientsOther CalibrationsCalibration FileContact PersonDocumentation TitleCalibration MethodCalibration NotesOther InformationDatalogger Entry DateResponsivity [45] [Corr]Calibration Factor [45] [Corr]Correction MethodResponsivity Bins Average [2 deg]
1S08029PREDE (Tokyo, Japan)05/19/200811/19/2008PREDEPGS-100Direct Normal SpectroradiometerOPTIC05/19/2008[100_2008_05.cal.zip]PGS-100 M#S08029 Calibration Constant.xlscal files available online
2S08029NREL/Optics Lab08/06/200802/06/2009PREDEPGS-100Direct Normal SpectroradiometerOPTIC05/19/2008[100_2008_08.rpt.zip][100_2008_08.cal.zip]Afshin AndreasPGS-100 Calibration ReportOutdoors vs LI-1800 PRS-158 w/Direct Tubecal files and report onlineReceived OUT of Tolerance08/06/2008
3S08029NREL/Optics Lab04/08/200910/08/2009PREDEPGS-100Direct Normal SpectroradiometerOPTIC05/19/2008[100_2009_04.rpt.zip][100_2009_04.cal.zip]Afshin AndreasPGS-100 Calibration ReportOutdoors vs LI-1800 PRS-158 w/Direct Tubecal files and report onlineIN Tolerance04/10/2009
4S08029NREL/Optics Lab11/17/200905/17/2010PREDEPGS-100Direct Normal SpectroradiometerOPTIC05/19/2008[100_2009_11.rpt.zip][100_2009_11.cal.zip]Afshin AndreasPGS-100 Calibration ReportOutdoors vs LI-1800 PRS-158 w/Direct Tubecal files and report online; shutter solenoid replaced at factoryIN Tolerance11/18/2009