R# | Serial Number | Calibration Facility | Calibration Date | Date for next Calibration | Responsivity [45] | Calibration Factor [45] | Instrument::Make | Instrument::Model | Instrument::Description | Instrument::Program | Instrument::Factory Calibration | Instrument::Date of Factory Calibration | Responsivity Bins [2 deg] | Responsivity Function Coefficients | Other Calibrations | Calibration File | Contact Person | Documentation Title | Calibration Method | Calibration Notes | Other Information | Datalogger Entry Date | Responsivity [45] [Corr] | Calibration Factor [45] [Corr] | Correction Method | Responsivity Bins Average [2 deg] |
1 | Q20555 | LI-Cor | 05/26/1995 | 05/26/1997 | CF = 319.62 µmol/s/m²/mV | CF = 193.05 µmol/s/m²/µA | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | | | | | Cal factor (per mV) calculated using 604 Ohm resistor in parallel | Photosynthetically Active Radiation (PAR) | SRRL BMS UV Radiometer | | | | | |
2 | Q20555 | NREL | 03/28/2006 | 03/28/2007 | CF = 328.9 µmol/s/m²/mV | | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | | | Afshin | | vs BMS Downwelling Quantum Sensor 11:26 to 13:03 MST | Photosynthetically Active Radiation (PAR) | SRRL BMS UV Radiometer | 03/28/2006 | | | | |
3 | Q20555 | NREL | 04/14/2008 | 04/14/2010 | CF = 306.5 µmol/s/m²/mV | | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | [BMS_GloPAR_2008_04.pdf] | | Afshin Andreas | SRRL Quantum Calibration Report | 400-700nm vs LI-1800 Photosynthetically Active Radiation (PAR) | Calibrated as Global Horiz (used as Upwelling) | SRRL BMS UV Radiometer | 04/16/2008 | | | | |
4 | Q20555 | NREL | 09/09/2010 | 09/09/2011 | CF = 336.61 µmol/s/m²/mV | | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | [BMS_GloPAR_2010_09.pdf] | | Afshin Andreas | SRRL Quantum Calibration Report | 400-700nm vs LI-1800 Photosynthetically Active Radiation (PAR) | Calibrated as Global Horiz (used as Upwelling) | SRRL BMS UV Radiometer | 09/10/2010 | | | | |
5 | Q20555 | NREL | 04/26/2013 | 04/26/2015 | CF = 330.10 µmol/s/m²/mV | | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | [BMS_GloPAR_2013_04.pdf] | | Afshin Andreas | SRRL Quantum Calibration Report | 400-700nm vs LI-1800 PRS-456 Photosynthetically Active Radiation (PAR) | Calibrated as Global Horiz (used as Upwelling) | SRRL BMS UV Radiometer | 06/10/2013 | | | | |
6 | Q20555 | SRRL | 06/14/2016 | 06/14/2018 | 2.928 uV/umol/s*m^2 | 341.58 umol/s/m^2/mV | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | | | Andreas | | vs Global EKO MS-7XX outdoors at noon | | SRRL BMS UV Radiometer | 06/15/2016 | | | | |
7 | Q20555 | SRRL | 09/14/2018 | 09/14/2020 | 3.000 uV/umol/s*m^2 | 333.32 umol/s/m^2/mV | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | | | Andreas | | vs Global EKO MS-711 outdoors at noon | Calibrated as Global Horiz (used as Upwelling) | SRRL BMS UV Radiometer | 09/14/2018 | | | | |
8 | Q20555 | SRRL | 09/29/2020 | 09/29/2022 | 2.893 uV/umol/s*m^2 | 345.71 umol/s/m^2/mV | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | | | Andreas | | vs Global EKO MS-711 outdoors at noon | Calibrated as Global Horiz (used as Upwelling) | SRRL BMS UV Radiometer | 09/29/2020 | | | | |
9 | Q20555 | SRRL | 05/13/2022 | 05/13/2024 | | 351.63 umol/s/m^2/mV | Li-Cor | LI-190 | Quantum (PAR) Sensor | NREL/Program | | 05/26/1995 | | | | | Andreas | | vs Global EKO MS-711 PAR outdoor 11:49 to 12:23 MST | Calibrated as Global Horiz (used as Upwelling) | SRRL BMS UV Radiometer | | | | | |