R# | Serial Number | Calibration Facility | Calibration Date | Date for next Calibration | Responsivity [45] | Calibration Factor [45] | Instrument::Make | Instrument::Model | Instrument::Description | Instrument::Program | Instrument::Factory Calibration | Instrument::Date of Factory Calibration | Responsivity Bins [2 deg] | Responsivity Function Coefficients | Other Calibrations | Calibration File | Contact Person | Documentation Title | Calibration Method | Calibration Notes | Other Information | Datalogger Entry Date | Responsivity [45] [Corr] | Calibration Factor [45] [Corr] | Correction Method | Responsivity Bins Average [2 deg] |
1 | 23169 | Solar Light | 09/05/2017 | 09/05/2018 | CF = 0.250 V/(MED/Hr) = 4 MED/Hr/V | | Solar Light | 501A-UVBiometer | Ultraviolet Radiometer | NREL/Program | 0.250 [(V/(MED/Hr))] | 09/05/2017 | | | | | Bekim Abazoski | Certification of Calibration | | 290 to 400nm (Erythema) | SRRL BMS UV Radiometer | 07/26/2018 | | | | |
2 | 23169 | NREL | 07/31/2018 | 07/31/2019 | CF = 1.591 W/m²/V | CF(erythema) = 3.770 MED/Hr/V = 8.792 Idx/V | Solar Light | 501A-UVBiometer | Ultraviolet Radiometer | NREL/Program | 0.250 [(V/(MED/Hr))] | 09/05/2017 | | | | | Afshin Andreas | | vs OL756/S9011; 280-315nm for UVB, 280-400nm for Erythema | CIE 1987 Erythema Action Spectrum | SRRL BMS UV Radiometer | 08/02/2018 | | | | |
3 | 23169 | NREL | 04/30/2022 | 04/30/2023 | CF = 1.625 W/m²/V (280-315nm) | CF = 0.2115 W/m²/V = 3.628 MED/Hr/V = 8.461 Index/V (erythema) | Solar Light | 501A-UVBiometer | Ultraviolet Radiometer | NREL/Program | 0.250 [(V/(MED/Hr))] | 09/05/2017 | | | | | Afshin Andreas | | vs OL756 Outdoors at solar noon | Uncert = ±6% (outdoors Z=24) | SRRL BMS UV Radiometer | 05/03/2022 | | | | |
4 | 23169 | PMOD | 07/12/2022 | | C = 0.2019 W/m²/V @ Z = 40 (erythema) | CF = 0.2118 W/m²/V @ Z = 25 (erythema) | Solar Light | 501A-UVBiometer | Ultraviolet Radiometer | NREL/Program | 0.250 [(V/(MED/Hr))] | 09/05/2017 | | | | | Afshin Andreas | 2022-2551-1-1 Calibration Certificate | | Uncert = ±3.2% | SRRL BMS UV Radiometer | | | | E(erythema) = (S-Sdark) * C * f(Z,Ozone) * Coscor(Z) | |