MIDC/Data Quality Statements
Total Records: 33, displaying 26-33
[Prev Records]
R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
26BMS0212-2Global TSP-70009/26/200800:00:0003/04/200914:00:00There is an occasional drastic noise problem with the SRRL BMS TSP-700, which started showing up 9/26/2008. Every screw between the data logger and instrument was retightened, this includes the patch panel and power supply box, which seemed to solve the problem.
27BMS0212-2Global TSP-70005/29/200800:00:0006/09/200813:00:00The signal from the TSP-700 would occasionally drop low during this time frame. A screw was tightened in the power supply box, which seemed to fix the problem.
28BMS0212-2Global TSP-70012/22/200708:41:0012/22/200711:24:00Due to a scheduled SRRL/Mesa Top power outage on 12/22/2007, the TSP-1, TSP-700, All Global UVA and UVB, and the TWC Photometer were without power, therefore data is either bad or bogus.
29BMS0212-2Global TSP-70005/25/200610:50:0005/25/200611:10:00Data from the Yankee instrument is bad or bogus due to power shutdown for installation of new SRRL equipment lift.
30BMS0212-2Global TSP-70011/18/200508:20:0011/18/200512:06:00The SRRL BMS Yankee TSP-700 was down for power supply upgrade. The old Condor HAA15-0.8-A+ (+/-12 V @ 1 Amp) was replaced with a new Condor HB12-1.7-A+ (+12 V @ 1.7 Amp). The new higher current power supply can now restart the TSP-700 after a power failure. Note: The -12 V output on the old power supply was not needed (Yankee had designed it as a universal power supply for all their instruments).
31BMS0212-2Global TSP-70011/04/200523:14:0011/09/200508:42:00The SRRL BMS Yankee TSP-700 has been down due to a design flaw or fault with the Yankee power supply, caused by a power outage on 11/4/2005 at 23:14 MST. The power supply is unable to power up both the ventilator and the heater simultaneously (even though it was designed to). Data during this time is bogus.
32BMS0212-2Global TSP-70007/23/200505:33:0007/25/200507:18:00The SRRL BMS Yankee TSP-700 has been down due to a design flaw or fault with the Yankee power supply, caused by a scheduled NREL power outage on 7/23/2005 at 5:33 MST. The power supply is unable to power up both the ventilator and the heater simultaneously (even though it was designed to). Data during this time is bogus.
33BMS0212-2Global TSP-70002/16/200509:02:0002/17/200508:51:00The SRRL BMS Yankee TSP-700 has been down due to a design flaw or fault with the Yankee power supply, caused by a scheduled SRRL power outage on 2/16/2005 at 9:02 MST. The power supply is unable to power up both the ventilator and the heater simultaneously (even though it was designed to). Data during this time is bogus.