R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
26 | BMS | 010034 | Diffuse CM22 | 07/10/2005 | 05:56:00 | 07/11/2005 | 08:00:00 | Sci-Tec Tracker #2 rotated the wrong direction and hung up on the signal cables, due to a momenatary power loss. The following instruments were UNSHADED: Diffuse CM22, Diffuse UVB-1 and the CG4. |
27 | BMS | | Diffuse CM22 | 05/02/2003 | 13:45:00 | 07/03/2003 | 08:30:00 | On May 2, 2003, the shade on the CM22 and PSP was removed during BORCAL, which caused the diffuse instruments to measure global. However on May 5, 2003, it was decided not to corrupt the baseline diffuse measurements with "global data", so the instruments were disconnected and/or shorted at the baseline datalogger connection. The instruments were later returned to normal service on July 3, 2003. All data during this time is bad or bogus. |
28 | BMS | 010046 | Diffuse CM22 | 08/24/2002 | 09:15:00 | 09/02/2002 | 17:30:00 | The shading ball on the CM22 was removed to allow the instrument to be calibrated in BORCAL 2002-04. Affected times (MST) were: 8-24-02 9:15 to 14:30, 8-25-02 10:00 to 15:00, 8-26-02 7:15 to 15:00, 8-27-02 10:15 to 12:45, 8-30-02 5:30 to 13:35, 9-2-02 10:00 to 17:30. |