December 31, 2014
- 15:01 MST: Data for CH1 has been repaired on Dec 24 through 27, see data quality statement for more details. -Afshin Andreas
- 10:43 MST: See BMS data quality statements for recent problems with Direct CH1, Direct TUVR, Direct CUVB2, and Global CUV4. -Afshin Andreas
- 16:24 MST: There is a possible problem with the SRRL BMS Downwelling PIR which started yesterday at 10:00 MST, which is causing a possible 10-15 W/m^2 error. Problem could be due to possible water infiltration into temporary J-box which was not oriented in the proper position to shed water. All temporary J-box have been re-oriented correctly, however connections for PIR may need to dry out. -Afshin Andreas
- 14:09 MST: The SRRL BMS HMP45C measuring Deck Dry Bulb Temp and Relative Humidity was replaced with a HMP60 at 10:09 MST yesterday. -Afshin Andreas
- 16:53 MST: Starting next week the singal cabling, junction boxes, and patch panel at SRRL will be replaced, this affects the BORCAL and BMS. Therefore a temporary data acquisition system was installed to support BMS during the upgrade which could take up to 2 months depending on weather. The temporary data acquisition system went online at 10:09 MST. The temporary system will only collect data from primary measurements, the following measurements will be unavailable during this time:
Global SP-110
Global UVB-1
Direct NIP (unit 2)
Shadowband PSP
Zebra PSP
All UV case temperatures
All BMS Research Channels
TWC Photometer/AOD
DataLab Temp/RH
Global PSP (unventilated), note this will be permanently removed from service.
-Afshin Andreas
- 15:37 MST: The Prede POM-02 Photometer at SolarTAC was permanently removed from service on October 29, 2014. -Afshin Andreas
- 12:28 MST: Intermittent problem with SRRL LI-1800 global spectroradiometer was caused by faulty serial communication cable. The cable has been replaced with a temporary cable which seems to have resolved the issue. -Afshin Andreas
- 15:55 MST: The SRRL BMS Upwelling PIR was swapped with calibrated sensor at 15:29 MST. The new sensor was temporarily installed on the radiometer tower CR3000 (instead of Deck CR3000) in preparation for deck cable upgrades later this month. -Afshin Andreas
- 09:24 MST: Data collection to the SRRL LI-1800 global spectroradiometer has been intermittent since Nov 9, instrument is old and may be failing. -Afshin Andreas
- 15:35 MST: The SRRL BMS HMP45C measuring Tower Dry Bulb Temp and Relative Humidity was replaced with a HMP60 at 10:41 MST today. -Afshin Andreas
- 15:34 MST: The data logger at the SRRL BMS Radiometer tower was upgraded from a CR10X-TD to a CR3000 between 10:40 and 11:45 MST today. The following sensors will be bogus during that time: Global CM3, Quantum LI-190, and Photometric LI-210; Upwelling CM3, CG3, PSP, and LI-200; Downwelling CG3; Global 40-S/90-N/90-E/90-W PSP & LI-200; Snow Depth; EFM; WS & WD @ 6ft. -Afshin Andreas
- 17:51 MST: The SRRL Direct Normal Quantum sensor was calibrated on November 6, 2014 outdoors against the NIST traceable LI-1800 PRS-102 spectroradiometer. The new calibration factor was applied today @ 17:48 MST. -Afshin Andreas
- 13:49 MST: The SRRL PGS-100 spectroradiometer was calibrated on November 6, 2014 outdoors against the NIST traceable LI-1800 PRS-102 spectroradiometer. The new calibration file was applied to data starting at 13:45 MST today. -Afshin Andreas
- 10:20 MST: The SRRL BMS LICOR LI-1800 Global Spectroradiometer was removed from service on 4-Nov-2014 and calibrated against NIST FEL F680 Lamp Standard. Instrument was within 2-3% of NIST values, instrument was recalibrated and returned to service. Unfortunately due to a problem with the serial communications cable, the data collection did not resume until sunrise today. -Afshin Andreas
- 14:26 MST: The SRRL BMS Global Normal CM21 was swapped out with a sensor calibrated in BORCAL 2014-02 between 8:50 and 9:15 today. Calibration factor was updated in logger at 9:05 MST today. Note: Moisture was observed in dome on Sept 12, hence the instrument was swapped out for this reason as well. -Afshin Andreas
- 07:31 MST: The SRRL ATI has been malfunctioning since 06/08/2014, data should not be used. -Afshin Andreas
- 15:49 MST: The filter correction factor for the SRRL BMS Global PSP RG780 was recomputed and applied at 15:28 MST today using RG780 filter transmission spectra vs clear dome transmission spectra that were measured on 6/20/2014. -Afshin Andreas
- 15:47 MST: The SRRL BMS Global TSP-1 and Diffuse CUV4 were permanently removed from service @ 14:58 MST today, values report missing (-99999). -Afshin Andreas
- 10:38 MST: The SRRL BMS Setra Pressure Sensor was swapped out with a calibrated sensor at 10:11 MST today. -Afshin Andreas
- 16:15 MST: The Excel® spreadsheet to help the user determine instrument measurement uncertainty has been updated. -Afshin Andreas
- 13:32 MST: The temperature, wind speed, and wind direction sensors at NWTC M2 were replaced with calibrated instruments on August 13, 2014. The data logger program was updated today @ 13:24 MST with new calibration constants for the wind speed and temperature sensors. -Afshin Andreas
- 07:06 MST: January 2013 through December 2013 Monthly Data Sets & Reports/Plots for SRRL BMS have been generated. -Afshin Andreas
- 08:18 MST: The (2) Eppley Ventilators for the SRRL BMS Global PSP and PSP RG780 were upgraded from AC to DC fans between 10:30 and 15:00 MST on July 3. -Afshin Andreas
- 13:28 MST: The SRRL BMS Global SE-POA LI-200 was swapped out with a sensor calibrated in BORCAL 2013-03 today at 13:21 MST (logger program was updated at same time). -Afshin Andreas
- 07:12 MST: The (3) Eppley Ventilators mounted on the SRRL BMS Sci-Tec tracker were upgraded from AC to DC fans between 12:30 and 17:00 MST yesterday. This applies to the Diffuse 8-48, Diffuse PSP, and Downwelling PIR. During this time data from the PSP is bogus, since the 8-48 and PIR were rotated to the shaded PSP position during upgrade to help maintain shading of 8-48 and PIR. The (2) ventilators for the Global PSP and RG780 PSP will be upgraded later this week. -Afshin Andreas
- 14:01 MST: The following SRRL BMS instruments were swapped with calibrated instruments between approx. 12:26 and 12:55 MST: Diffuse 8-48, CM22, PSP (calibrated in BORCAL 2014-02), RG780 NIP&PSP and Diffuse Shadowband&Zebra (calibrated in borcal 2013-02 and 2013-03), and Downwelling PIR&CG4 (calibrated 28-May-2013). Calibration factors were update in program at 12:29 MST today, along with the Global CM6b (calibrated in BORCAL 2014-02). -Afshin Andreas
- 12:29 MST: As of 13-June-2014 at 11:22 MST, the obsolete responsivity function (separate AM&PM functions) used for the SolarTAC CMP22 was replaced with spline fit to the AM/PM averaged 2-degree responsivity data. All computations are now done in the logger with no post processing, this includes zenith, azimuth, airmass, zenith corrected global, wind chill, and dew point. -Afshin Andreas
- 12:27 MST: SolarTAC CMP22, CH1, and 8-48 instruments were swapped out with calibrated sensors today between approx. 11:00 and 11:30 MST. Cal factors were updated in logger program at 11:22 MST. -Afshin Andreas
- 14:00 MST: SRRL TSI-880 Total Sky Imager shadowband failed on June 5, 2014. Repairs are TBD. -Afshin Andreas
- 08:22 MST: All SRRL BMS UV instruments were calibrated on 05-28-2014, this includes: UV-S-B-T, CUVB1, MS210W, UVB-1, 501A, TUVR (both), CUV4 (global only), UV-S-A-T, CUVA1, CUVA2, CUVB2. Logger was updated with new calibration factors yesterday @ 15:00 MST. -Afshin Andreas
- 08:19 MST: The following SRRL BMS instruments were calibrated in BORCAL 2014-01: Global Horizontals: CM22, PSP (ventilated), PSP, TSP-700, TSP-1, SP-Lite, LI-200, SP-110, SPN1; Direct Normals: CH1, MS-56, NIP #2, NIP #1, LI-201; Diffuse SPN1. Logger program was updated at 15:00 MST yesterday with new calibration factors (15:04 MST for PSPs). -Afshin Andreas
- 09:26 MST: MIDC Bug Fix: Corrected gnuplot issue which results in "Error: Plot output is empty" message when data on one y-axis is all missing, however the opposite y-axis has valid data. Data for valid data is now displayed. -Afshin Andreas
- 09:24 MST: New MIDC feature: After bringing up the maintenance records for a measurement, the dates now have a link which will allow you to see all records for that date. -Afshin Andreas
- 07:08 MST: There was a scheduled power outage at SRRL on 5/20/2014 @ 19:30 through 21:12 MST for the installation of a generator transfer switch. Since this was at night there was no data loss (except for the Atmospheric Electric Field Monitor). -Afshin Andreas
- 07:21 MST: SRRL BMS data on December 21, 2013 from 10:30 to 16:37 MST was corrected due to NREL power outage. The following instruments were corrected:
- Direct CH1 was replaced with data from SRRL RSR2 Direct Normal
- Diffuse 8-48 was calculated using BMS Global CM22 and SRRL RSR2 Direct Normal
- Diffuse CM22 was replaced with new BMS Diffuse 8-48 data (all day long).
- 08:27 MST: The SRRL BMS Data Lab Temp/RH probe was removed from service on April 23 @ 8:30 MST for calibration. -Afshin Andreas
- 09:42 MST: There is a problem with the SolarTAC tracker since Feb 4 (gears are grinding), therefore direct is zero and diffuse is unshaded. -Afshin Andreas
- 14:13 MST: An EKO MS-56 pyheliometer has been added to the SRRL Baseline Measurement System (BMS) data stream. Data is available since January 6, 2014. -Afshin Andreas
- 14:11 MST: SRRL Aerosol Optical Depth (AOD) data streams have been added to MIDC. Data is processed by the European Sky Radiometers (ESR) Network and is available since June 11, 2013. -Afshin Andreas