MIDC/Data Quality Statements
Total Records: 6, displaying 1-6

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSUpwelling LI-19005/13/202211:44:0005/13/202212:25:00The LICOR Quantum sensor was facing toward the sun during this time so that it could be calibrated. Data should not be used during this time.
2BMSUpwelling LI-19005/12/202110:20:0005/12/202112:13:00The relay multiplexer on the SRRL BMS radiometer tower had a problem (old age/possible moisture) that caused low voltage measurements on the main CR3000 wiring panel to read bogus, this includes all 40-South measurements and all upwelling shortwave measurements (except upwelling CMP11).
3BMSUpwelling LI-19009/29/202011:43:0009/29/202012:18:00The LICOR Quantum sensor was facing toward the sun during this time so that it could be calibrated. Data should not be used during this time.
4BMSUpwelling LI-19009/14/201811:33:0009/14/201812:28:00The LICOR Quantum sensor was facing toward the sun during this time so that it could be calibrated. Data should not be used during this time.
5BMSUpwelling LI-19006/14/201610:47:0006/14/201614:50:00The LICOR Quantum sensor was facing toward the sun during this time so that it could be calibrated. Data should not be used during this time.
6BMSQ20558Upwelling LI-19003/22/200615:00:0003/28/200614:32:00The LICOR Quantum sensor is reading very low (and negative when irradiance levels are low enough, near sunrise/set). Sensor was replaced with S/N Q20555.