R# | Site | Serial Number | Measurement | Start Date | Start Time | End Date | End Time | Data Quality Statement |
1 | BMS | | Diffuse Shadowband | 05/13/2020 | 12:19:00 | 12/15/2021 | 19:31:00 | DATA CORRECTED on 12/17/2021: Irradiances were increased by 0.675% due to revised BORCAL report for all BMS shortwave radiometers calibrated in 2020 and 2021. |
2 | BMS | 34722 | Diffuse Shadowband | 01/29/2019 | 07:30:00 | 02/04/2019 | 08:20:00 | The shadowband was out of adjustment during this time, the diffuse 8-48 shadowband will read high under some conditions. |
3 | BMS | | Diffuse Shadowband | 03/29/2018 | 06:00:00 | 04/02/2018 | 08:20:00 | The shadowband was slightly out of adjustment during this time, the diffuse 8-48 shadowband will read high under some conditions. |
4 | BMS | | Diffuse Shadowband | 08/13/2017 | 06:00:00 | 08/15/2017 | 18:00:00 | The shadowband was slightly out of adjustment during this time, the diffuse shadowband will read high under some conditions. |
5 | BMS | 31395F3 | Diffuse Shadowband | 11/29/2012 | 20:00:00 | 12/04/2012 | 13:52:00 | Signal was erratic or missing during this time. Screws inside patch panel were retightened which seemed to fix the problem. |
6 | BMS | | Diffuse Shadowband | 03/23/2012 | 13:10:00 | 03/30/2012 | 11:44:00 | Signal was erratic/very high or missing during this time. The screws on the patch panel were retightened. |
7 | BMS | | Diffuse Shadowband | 03/28/2011 | 00:00:00 | 03/28/2011 | 14:39:00 | Signal was low in the morning before 8:30 (may have been snow/frost), however the signal become large and strange after 13:00. Signals were disconnected and troubleshot on the deck and the data logger, no problem was found. Signal was back to normal again at 14:39. |
8 | BMS | 31399F3 | Diffuse Shadowband | 09/28/2009 | 06:00:00 | 10/05/2009 | 07:53:00 | Signal was high and noisy during this time. A screw on the patch panel was loose and was tightened. |
9 | BMS | 31399F3 | Diffuse Shadowband | 08/10/2009 | 09:29:00 | 08/11/2009 | 07:16:00 | Signal was erratic or missing during this time. Screws on the multiplexer were tightened. |