MIDC/Data Quality Statements
Total Records: 11, displaying 1-11

R#SiteSerial NumberMeasurementStart DateStart TimeEnd DateEnd TimeData Quality Statement
1BMSDiffuse CUV412/21/201307:00:0012/21/201316:22:00Due to a scheduled NREL power outage, to upgrade campus electrical infrastructure, The following BMS instruments were not pointing at the sun during this time: Direct NIP #1, LI-201, LI-190, NIP RG780, TWC; and the following instruments were un-shaded: Diffuse CM22, CUV4, CG4. The Middle Sci-Tec tracker for these instruments were temporarily put on a secondary UPS which failed immediately.
2BMSDiffuse CUV403/02/201313:06:0003/02/201318:24:00Due to a scheduled NREL power outage, to upgrade campus power capacity, The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, NIP #1, LI-201, NIP RG780, CUVA2, CUVB2, TWC; and the following instruments were un-shaded: Diffuse 8-48, PSP, CM22, CUV4, PIR, CG4. The Sci-Tec tracker for these insruments are protected by UPS, however the battery depleted at 13:06 MST (the power outage started at 8:06 MST)
3BMSDiffuse CUV405/26/201206:15:0005/26/201207:12:00Due to a planned power outage for SRRL electrical work, the following BMS instruments were not following the sun: NIP #1, RG780 NIP, LI-201, TWC Photometer and the following instruments were not shaded CM22, CUV4, and CG4.
4BMSDiffuse CUV409/08/201108:23:0009/08/201109:38:00Due to a planned power outage for SRRL construction activities, the following BMS instruments were not following the sun: NIP #1, RG780 NIP, LI-201, TWC Photometer and the following instruments were not shaded CM22, CUV4, and CG4.
5BMSDiffuse CUV405/07/201105:55:0005/09/201107:58:00There was a planned power outage for SRRL construction activities that started at 5:18 MST. Unfortunately, the backup generator provided to SRRL was providing bad power to the UPS powering the Sci-Tec trackers, therefore the UPS failed at 5:55 MST. Unfortunately, when normal AC power was restored at 11:45 MST, the Sci-Tec trackers had lost all their configuration due to UPS cycling on and off during the outage, therefore were not tracking. The tracker settings were restored on May 9 at 7:58 MST. The following BMS instruments were not pointing at the sun during this time: Direct CH1, NIP #2, NIP #1, LI-201, NIP RG780, CUVA2, CUVB2, TWC; and the following instruments were un-shaded: Diffuse 8-48, PSP, CM22, CUV4, PIR, CG4
6BMSDiffuse CUV403/18/201100:00:0003/18/201107:57:00The Sci-Tec #2 tracker was jammed on the shading plate. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201, TWC Photometer.
7BMS080018Diffuse CUV410/30/201000:00:0011/01/201011:27:00The Sci-Tec #2 tracker was apparently not working during this time. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201.
8BMSDiffuse CUV412/15/200909:40:0012/15/200910:00:00The Sci-Tec #2 tracker collided with another empty Sci-Tec tracker. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201.
9BMSDiffuse CUV412/09/200912:11:0012/09/200914:22:00Due to a power outage at NREL, followed by bad power, the SRRL tracker UPS shutdown. The following BMS instruments were not pointing at the sun: Direct CH1, NIP1, NIP2, NIP-RG780, LI201, TWC500, CUV4; and the following instruments were unshaded: 8-48, CM22, PSP, CUV4, PIR, CG4.
10BMS080018Diffuse CUV407/30/200912:45:0007/30/200923:59:00The Sci-Tec #2 tracker was apparently not working during this time. The following instruments were unshaded: CM22, CUV4, CG4; and the following instruments were not pointing at the sun: NIP #1, NIP RG780, LI-201.
11BMSDiffuse CUV411/15/200814:02:0011/15/200814:32:00Due to a planned power outage at SRRL, the following BMS instruments were unshaded: Diffuse 8-48, CM22, PSP, CUV4, DW-PIR, DW-CG4.