R# | Serial Number | Calibration Facility | Calibration Date | Date for next Calibration | Responsivity [45] | Calibration Factor [45] | Instrument::Make | Instrument::Model | Instrument::Description | Instrument::Program | Instrument::Factory Calibration | Instrument::Date of Factory Calibration | Responsivity Bins [2 deg] | Responsivity Function Coefficients | Other Calibrations | Calibration File | Contact Person | Documentation Title | Calibration Method | Calibration Notes | Other Information | Datalogger Entry Date | Responsivity [45] [Corr] | Calibration Factor [45] [Corr] | Correction Method | Responsivity Bins Average [2 deg] |
1 | S92096.04 | EKO | 05/13/1993 | | 200 | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | | | | Calibration Certificate | | | | | | | | |
2 | S92096.04 | NREL | 09/07/2000 | 09/07/2001 | CF = 0.3489 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2000_09b.zip] | | Afshin Andreas | SRRL BMS UVB Calibration Report B | 280-315nm vs OL754 | CF @ 305nm peak = 0.0121 W/m²/nm/mV | SRRL BMS UV Radiometer | 11/01/2000 | | | | |
3 | S92096.04 | NREL | 08/17/2001 | 08/17/2002 | CF = 0.3832 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2001_08.zip] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL754 | | SRRL BMS UV Radiometer | 08/19/2001 | | | | |
4 | S92096.04 | NREL | 08/26/2002 | 08/26/2003 | CF = 0.4319 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2002_08.zip] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL754 | | SRRL BMS UV Radiometer | 08/26/2002 | | | | |
5 | S92096.04 | NREL | 10/29/2003 | 10/29/2004 | CF = 0.5425 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2003_10.zip] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL754 | | SRRL BMS UV Radiometer | 11/05/2003 | | | | |
6 | S92096.04 | NREL | 09/15/2004 | 09/15/2005 | CF = 0.6945 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2004_09.zip] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL754 | | SRRL BMS UV Radiometer | 10/18/2004 | | | | |
7 | S92096.04 | NREL | 08/29/2005 | 08/29/2006 | CF = 0.8723 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2005_08.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 09/20/2005 | | | | |
8 | S92096.04 | NREL | 09/12/2006 | 09/12/2007 | CF = 1.0785 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2006_09.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 09/13/2006 | | | | |
9 | S92096.04 | NREL | 10/26/2007 | 10/26/2008 | CF = 1.3367 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2007_10.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 03/18/2008 | | | | |
10 | S92096.04 | NREL | 10/26/2009 | 10/26/2010 | CF = 1.4805 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2009_10.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 10/30/2009 | | | | |
11 | S92096.04 | NREL | 05/31/2011 | 05/31/2012 | CF = 1.4766 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2011_05.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 06/17/2011 | | | | |
12 | S92096.04 | NREL | 06/13/2012 | 06/13/2013 | CF = 1.5703 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2012_06.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 07/20/2012 | | | | |
13 | S92096.04 | NREL | 05/28/2014 | 05/28/2015 | CF = 1.651 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2012_06.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 06/03/2014 | | | | |
14 | S92096.04 | NREL | 08/05/2015 | 08/05/2016 | CF = 1.708 W/m²/mV | | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | [BMS_GloUVB_2012_06.pdf] | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 08/13/2015 | | | | |
15 | S92096.04 | NREL | 06/08/2017 | 06/08/2018 | CF = 1.886 W/m²/mV | CF = 18.88 W/m²/V (using 1 V output) | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | | | Afshin Andreas | SRRL BMS UVB Calibration Report | 280-315nm vs OL756 | 1V output calibrated on 11/13/2017 (99.9 times higher than 10mV output) | SRRL BMS UV Radiometer | 06/09/2017 | | | | |
16 | S92096.04 | NREL | 07/31/2018 | 07/31/2019 | | CF = 21.41 W/m²/V | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | | | Afshin Andreas | | 280-315nm vs OL756/S9011 | | SRRL BMS UV Radiometer | 08/02/2018 | | | | |
17 | S92096.04 | NREL | 07/10/2020 | 07/10/2021 | | CF = 21.40 W/m²/V | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | | | Afshin Andreas | | 280-315nm vs OL756 | | SRRL BMS UV Radiometer | 07/13/2020 | | | | |
18 | S92096.04 | NREL | 04/30/2022 | 04/30/2023 | | CF = 25.37 W/m²/V (280-315nm) | EKO | MS-210W | UVB Sensor | NREL/Program | 200 | 05/13/1993 | | | | | Afshin Andreas | | vs OL756 Outdoors at solar noon | Uncert = ±6% (outdoors Z=24) | SRRL BMS UV Radiometer | 05/05/2022 | | | | |