MIDC/AIM Calibration Database
Total Records: 9, displaying 1-9

R#Serial NumberCalibration FacilityCalibration DateDate for next CalibrationResponsivity [45]Calibration Factor [45]Instrument::MakeInstrument::ModelInstrument::DescriptionInstrument::ProgramInstrument::Factory CalibrationInstrument::Date of Factory CalibrationResponsivity Bins [2 deg]Responsivity Function CoefficientsOther CalibrationsCalibration FileContact PersonDocumentation TitleCalibration MethodCalibration NotesOther InformationDatalogger Entry DateResponsivity [45] [Corr]Calibration Factor [45] [Corr]Correction MethodResponsivity Bins Average [2 deg]
129238NREL/Optics Lab05/16/199605/16/19974.46 W/sm/mVEppleyTUVRUltraviolet RadiometerOPTICDaryl Myerssee paper fileuncert=+/-15%
229238NREL/Optics Lab08/28/199608/28/1997Cf=4.461 W/sm/mvEppleyTUVRUltraviolet RadiometerOPTICD Myers
329238NREL01/06/200001/06/20013.350 W/m^2/mVEppleyTUVRUltraviolet RadiometerOPTICAfshin AndreasCalibration ReportOutdoors using OL754
429238NREL/Optics Lab12/12/201304/12/2014CF = 5.5 W/m²/mVEppleyTUVRUltraviolet RadiometerOPTICAfshin Andreas280-400nm vs BMS Calculated Direct Total UV (Global & Diffuse CUV4) at noonTUVR w/ Custom Direct Tube & WindowSRRL BMS UV Radiometer12/12/2013
529238NREL05/28/201405/28/2015CF = 5.247 W/m²/mVEppleyTUVRUltraviolet RadiometerOPTIC[BMS_DirUVT_2012_06.pdf]Afshin AndreasSRRL BMS UV-Total Calibration Report280-400nm vs OL756 w/direct tubeTUVR w/ NREL custom direct tube & windowSRRL BMS UV Radiometer06/03/2014
629238NREL07/14/201607/14/2016CF = 5.092 W/m²/mVEppleyTUVRUltraviolet RadiometerOPTICAfshin AndreasSRRL BMS UV-Total Calibration Report280-400nm vs OL756 w/direct tubeTUVR w/ NREL custom direct tube & windowSRRL BMS UV Radiometer07/14/2016
729238NREL07/31/201807/31/2019CF = 5.099 W/m²/mVEppleyTUVRUltraviolet RadiometerOPTICAfshin Andreas280-400nm vs OL756 w/direct tubeTUVR w/ NREL custom direct tube & windowSRRL BMS UV Radiometer08/02/2018
829238NREL07/31/202007/31/2021CF = 4.962 W/m²/mVEppleyTUVRUltraviolet RadiometerOPTICAfshin Andreas280-400nm vs OL756 w/direct tubeTUVR w/ NREL custom direct tube & windowSRRL BMS UV Radiometer08/03/2020
929238NREL07/08/202207/08/2023CF = 4.898 W/m²/mV (280-400nm)EppleyTUVRUltraviolet RadiometerOPTICAfshin Andreasvs OL756 w/Direct Tube Outdoors Uncert = ±6% (outdoors)SRRL BMS UV Radiometer07/08/2022